Compound extraction from whole lung of rats with Precellys®24 Dual – 7 ml vial - Bertin Instruments

Other

Compound extraction from whole lung of rats with Precellys®24 Dual – 7 ml vial

Sources: Chiesi Farmaceutici, DMPK department – Italy

Context

The Company is focused on respiratory diseases. Our goal is the detection of chemicals entities as potential news drugs in biological fluids and tissues, as lung, bronchus and trachea after administration to animal. The sample preparation involves homogenization of tissues, most of the times, whole organ hence the trials with Precellys®24 Dual – 7 ml vial.

Results

The analyte levels in the sample are almost the same, so no difference involving the two homogenizing

techniques are detectable (See Figure 1).

Reproducibility is good in both cases, taking into account the individual variability.

img-03712-810-DU016

Few residual cartilaginous fragments (probably

bronchus/trachea) were observed in both methods (fewer amounts with Ultraturrax). The results show that there is no impact on the analyte recovery from matrix.

With both methods, temperature within the sample after processing could reach 70˚C depending on the sample and protocols. Considering the stability of the analyte and the short time of exposure to this temperature (sample were preserved in ice before and after processing), sample degradation was not significant, but a refrigerating device could be useful in different situation.

Ultraturrax is fastidious and time consuming with cleaning steps and potential contamination between samples. The spent time is at least 5 times less using Precellys®24 Dual.

ANY QUESTIONS? CONTACT AN EXPERT

Please fill in this form if you wish to contact our sales team or receive further information.

My Account